Cable connections
Connect the plug of the cable to the socket (10) at the rear of the chamber.
The socket (10) is marked with a sticker:
Socket (10)
(14)
(13)
Figure 38: Connections at the chamber rear
(3)
Hose connections
Plug the freshwater hose into the hose connection (16) above the freshwater can and secure it with
a hose clamp. You can use a part of the standard supplied water hose.
Screw the hose nozzle (brass) to the free edge of the hose and screw it directly onto the freshwater
connection “IN” (13) at the rear of the chamber.
When the freshwater can is empty, the message
Freshwater supply
will be displayed on the controller
(chap. 12.1.4), the buzzer sounds, and the humidification module turns off. After acknowledging the alarm,
the humidification module tries to fill up and start operating.
To guarantee humidification during 24 hours even at high humidity set-points with manual wa-
ter supply, we recommend filling the freshwater can (option) at the end of each day.
Summary of Contents for KBF P 240-UL
Page 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Page 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Page 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Page 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Page 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Page 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Page 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Page 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Page 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...