With temperatures below +3 °C /
37.4 °F
, water must be completely removed from the humidifying sys-
tem.
CAUTION
Storage below +3 °C / 37.4 °F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the chamber.
Contact BINDER Service before any transportation below +3 °C /
37.4 °F
.
Permissible ambient humidity:
max. 70 % r.H., non-condensing
CAUTION
Condensation by excess humidity.
Danger of corrosion on the housing after operating at humidity values > 70 % r.H. for
a long period.
Dry the chamber completely before shut-down:
•
Set the humidity to 0 % r.H. To enable dehumidification, the humidifying and dehu-
midifying system must be activated (deactivated operation line “Humidity off”, chap.
8.4 and setting “Control on”, chap. 6.3).
•
Set the temperature set point to 60 °C /
140 °F
for approx. 2 hours (Manual mode).
•
Only then, shut down the chamber at the main power switch (1) and close the tap of
the water supply.
When after storage in a cold location you transfer the chamber to its warmer installation site, condensa-
tion may form. Before start-up, wait at least one hour until the chamber has attained ambient temperature
and is completely dry.
3.4
Location of installation and ambient conditions
Set up the chamber on a flat, even surface, and in a well-ventilated, dry location and align it using a spirit
level. The site of installation must be capable of supporting the chamber’s weight (see technical data,
chap. 25.4). The chambers are designed for setting up inside a building (indoor use).
CAUTION
Danger of overheating.
Damage to the chamber.
∅
Do NOT set up chamber in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
•
Permissible ambient temperature range during operation
: +18 °C /
64.4 °F
to +32 °C /
89.6 °F
. At
elevated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +22 °C +/- 3 °C /
71.6 °F ± 5.4 °F
to which the specified technical data relate. Devi-
ations from the indicated data are possible for other ambient conditions. Lower values of the
temperature range indicated in the technical data are valid at an ambient temperature of max.
25 °C
/ 77 °F
.
With each degree of ambient temperature >25 °C /
77 °F
, the refrigeration power decreases by
1.5 K.
Summary of Contents for KBF P 240-UL
Page 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Page 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Page 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Page 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Page 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Page 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Page 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Page 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Page 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...