KBF P + KBF LQC + KBWF (E6) 11/2016
page 166/189
Fault description
Possible cause
Required measures
Controller
(continued)
Program keeps the last program
setpoint constant while in setting
“ramp”.
Program line with setting “ramp”
is incomplete.
When programming with setting
“ramp”, define the end value of
the desired cycle by adding an
additional section with a section
time of at least one second.
Ramp temperature transitions are
only realized as steps.
Setting “step” has been selected. Select setting “ramp”.
Humidity alarm when operating
without water connection.
Humidity control turned on.
Turn off humidity control (chap.
6.3).
Acknowledging the alarm does
not cancel the alarm state.
Cause of alarm persists.
Remove cause of alarm. If the
alarm state continues, contact
BINDER service.
Alarm message:
- - - -
or <-<-<
or
>->->
Sensor rupture between sensor
and controller or Pt 100 sensor
defective.
Contact BINDER service.
Short-circuit.
Miscellaneous
Fluorescent tube does not illumi-
nate.
Defective fluorescent tube.
Replace the fluorescent tubes.
Impaired valve function of hose
burst protection.
Calcification.
Remove calcifications by citric
acid or acetic acid solutions
(chap. 4.3.4).
Have a plumber inspect the valve.
Only qualified service personnel authorized by BINDER must perform repair.
Repaired chambers must comply with the BINDER quality standards.
Summary of Contents for KBF P 240-UL
Page 173: ...KBF P KBF LQC KBWF E6 11 2016 page 173 189 25 7 Dimensions size 240...
Page 174: ...KBF P KBF LQC KBWF E6 11 2016 page 174 189 25 8 Dimensions size 720...
Page 176: ...KBF P KBF LQC KBWF E6 11 2016 page 176 189...
Page 177: ...KBF P KBF LQC KBWF E6 11 2016 page 177 189 26 2 EU Declaration of Conformity for KBF LQC...
Page 178: ...KBF P KBF LQC KBWF E6 11 2016 page 178 189...
Page 179: ...KBF P KBF LQC KBWF E6 11 2016 page 179 189 26 3 EU Declaration of Conformity for KBWF...
Page 180: ...KBF P KBF LQC KBWF E6 11 2016 page 180 189...
Page 182: ...KBF P KBF LQC KBWF E6 11 2016 page 182 189...
Page 183: ...KBF P KBF LQC KBWF E6 11 2016 page 183 189 27 Product registration...