SCHEDULE
Canadian Certificate Of Conformity No: FM16CA0144X
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA
T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail:
www.fmapprovals.com
F 348 (Mar 16)
Page 2 of 3
11. The marking of the equipment shall include:
Class I Division 1, Groups B, C, D
Class II, III, Division 1, Groups E, F, G
T6, Ta = -40°C to +65
T5, Ta = -40°C to +80°C
Type 4X
12.
Description of Equipment:
General
- The LB 480 Series Scintillation Measurement Equipment is used as part of a measuring system for
monitoring industrial processes. The equipment is used for continuously measuring the level or weight per unit
area, in tanks or bins, of liquid, granular, viscous or encrustation-forming media and for measuring conveyor belt
charges as well as the density of liquids, suspensions, slurries and bulk solids. The measurement principle is
based on the absorption of gamma rays. The radiation source is not part of the measuring equipment and is not
included in the product Approval.
Construction
- The LB 480 Series Scintillation Measurement Equipment consists of a scintillation detector with
associated electronics in an explosionproof/dust-ignitionproof enclosure with separate sensor and wiring
compartments. The sensor and wiring compartments are separated by NRTL listed cemented feedthroughs.
The sensor and housing (socket) are constructed of 304 or 316 stainless steel. The wiring compartment
contains four ½-inch NPT entries and contains a threaded cover. The socket attaches to the sensor by
cylindrical joint fastened with six bolts. The bolt holes terminate under the threaded cover inside the wiring
compartment) and therefore contain six threaded plugs to form a valid flamepath. The equipment is available
with an optional window which is cemented into the sensor.
Ratings
- The equipment is rated for use in an ambient temperature of -40°C to +65°C (or +80°C). The
equipment operates at 100-240 Vac (12 VA) or 24 Vdc (12 W).
LB 480-a-bA-cd-ef-0g0-h. Scintillation Measurement Equipment.
a = Sensor: 11, 12, 13, 14, 15, 16, 2A, 2B, 2E, 2F, 2I, 2J, 2K, 2L, 31, 32, 41, 42, 43 or 44.
b = Approval: F or G.
c = Signal output: 0 or 1.
d = Power supply: 1 or 2.
e = Collimator: 0, a, r, f or s.
f = Housing material: 1 or 3.
g = SIL: 0 or S.
h = Software: 0, S, T, U, V, W, L, M, N, O, D, E, F, G, H, J, P or K.
13.
Specific Conditions of Use:
The ambient temperature range and T-code rating for the equipment is as follows:
Ambient Temperature
T-Code
-40°C to +65°C
T6
-40°C to +80°C
T5
Summary of Contents for LB 480
Page 2: ......
Page 11: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 1 Safety Manual...
Page 12: ...54733 10BA2L 1 12 02 2017 Volume 1...
Page 22: ...54733 10BA2L 1 22 02 2017 4 Transport and Assembly Volume 1...
Page 56: ...54733BA26 1 56 02 2017 5 Explosion Protection Volume 1...
Page 57: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 57 Volume 1 5 Explosion Protection 1...
Page 58: ...54733BA26 1 58 02 2017 5 Explosion Protection Volume 1...
Page 59: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 59 Volume 1 5 Explosion Protection 1...
Page 60: ...54733BA26 1 60 02 2017 5 Explosion Protection Volume 1...
Page 61: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 61 Volume 1 5 Explosion Protection 1...
Page 62: ...54733BA26 1 62 02 2017 5 Explosion Protection Volume 1...
Page 63: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 63 Volume 1 5 Explosion Protection 1...
Page 64: ...54733BA26 1 64 02 2017 5 Explosion Protection Volume 1...
Page 66: ...54733BA26 1 66 02 2017 5 Explosion Protection Volume 1...
Page 67: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 67 Volume 1 5 Explosion Protection 1...
Page 68: ...54733BA26 1 68 02 2017 5 Explosion Protection Volume 1...
Page 69: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 69 Volume 1 5 Explosion Protection 1...
Page 70: ...54733BA26 1 70 02 2017 5 Explosion Protection Volume 1...
Page 71: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 71 Volume 1 5 Explosion Protection 1...
Page 72: ...54733BA26 1 72 02 2017 5 Explosion Protection Volume 1...
Page 73: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 73 Volume 1 5 Explosion Protection 1...
Page 108: ...54733 10BA2L 1 108 02 2017 7 Functional Safety Volume 1 7 13 Functional Safety Certificate...
Page 110: ...54733 10BA2L 1 110 02 2017 7 Functional Safety Volume 1...
Page 112: ...54733 10BA2L 1 112 02 2017 8 Visual Inspection Volume 1...
Page 128: ...54733 10BA2L 1 128 02 2017 9 Radiation Protection Volume 1...
Page 147: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 1 147 Volume 1 1 Notes...
Page 149: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 2 Installing SENSseries...
Page 150: ...54733 10BA2L 2 150 02 2017 Volume 2...
Page 170: ...54733 10BA2L 2 170 02 2017 1 System Description Volume 2...
Page 240: ...54733 10BA2L 2 240 02 2017 4 Repair Maintenance and Upkeep Volume 2...
Page 288: ...54733 10BA2L 2 288 02 2017 5 Technical Information Volume 2...
Page 297: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 2 297 Volume 2 2 Notes...
Page 299: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 3 Operation with HART Communicator...
Page 300: ...54733 10BA2L 3 300 02 2017 Volume 3...
Page 390: ...54733 10BA2L 3 390 02 2017 3 Getting Started via the HART Communicator Volume 3...
Page 410: ...54733 10BA2L 3 410 02 2017 6 Functional Processes Volume 3...
Page 428: ...54733 10BA2L 3 428 02 2017 7 Explanations Volume 3...
Page 440: ...54733 10BA2L 3 440 02 2017 8 Error Handling Volume 3...
Page 444: ...54733 10BA2L 3 444 02 2017 9 Setup Protocol Volume 3 Path Parameters Unit Standard SETUP...
Page 445: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 3 445 Volume 3 3 Notes...
Page 449: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 1 449 Volume 1 Notes...