SENSseries LB 480
BERTHOLD TECHNOLOGIES GmbH & Co. KG
1 – 77
Volume 1
5 Explosion Protection
1
Certificate:
70009819
Project:
70066628
Master Contract:
215040
Date Issued:
2016-03-07
DQD 507 Rev. 2012-05-22
Page 4
Input Ratings:
Variant
Type
Input Rating
Ax, Bx
LB 480-1x-xx-x2
LB 480-2x-xx-x2
100Vac-240Vac, 50/60Hz, 12VA
Ax, Bx
LB 480-1x-xx-x1
LB 480-2x-xx-x1
24Vdc, 12W
Ex
LB 480-3x-xx-x2
LB 480-4x-xx-x2
100Vac-240Vac, 50/60Hz, 12VA
Ex
LB 480-3x-xx-x1
LB 480-4x-xx-x1
24Vdc, 12W
Thermometer Circuit (PT100) Terminals 15 and 16
Type of protection Intrinsic Safety Ex ia IIC Maximum Values:
U
o
= 14.0 V; I
o
= 27.7 mA; P
o
= 97.0 mW; C
i
= 11 nF; L
i
= nil
Maximum permissible external values for common effective reactance’s (C
i
is not considered)
L
o
(mH)
C
o
(µF)
0.1
0.73
0.5
0.71
1.0
0.59
The RTD circuit is electrically connected to the internal supply circuit and the earth.
Open Collector Circuit (Terminal 11, 12)
Type of protection Intrinsic Safety Ex ia IIC Maximum Values:
U
i
= 15.0 V; I
i
= 26.6 mA; P
i
= 100.0 mW; C
i
= 11 nF; L
i
= nil
The open collector circuit is safely electrically isolated from earth and all other circuits.
Hart-current output (Source Mode) (Terminal 17, 18)
Type of protection Intrinsic Safety Ex ia IIC Maximum Values:
U
o
= 25.2 V; I
o
= 101.0 mA; P
o
= 635.0 mW;
Characteristic Linear: C
i
= 3.0 nF; L
i
= 20.0 µH
Maximum permissible external values for common
effective reactance’s (C
i
is not considered)
L
o
(mH)
C
o
(µF)
0.1
0.107
0.5
0.078
1.0
0.061
Single reactance’s to table A.2 and figure A.4 or A.6 of 60079
-11
Summary of Contents for LB 480
Page 2: ......
Page 11: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 1 Safety Manual...
Page 12: ...54733 10BA2L 1 12 02 2017 Volume 1...
Page 22: ...54733 10BA2L 1 22 02 2017 4 Transport and Assembly Volume 1...
Page 56: ...54733BA26 1 56 02 2017 5 Explosion Protection Volume 1...
Page 57: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 57 Volume 1 5 Explosion Protection 1...
Page 58: ...54733BA26 1 58 02 2017 5 Explosion Protection Volume 1...
Page 59: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 59 Volume 1 5 Explosion Protection 1...
Page 60: ...54733BA26 1 60 02 2017 5 Explosion Protection Volume 1...
Page 61: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 61 Volume 1 5 Explosion Protection 1...
Page 62: ...54733BA26 1 62 02 2017 5 Explosion Protection Volume 1...
Page 63: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 63 Volume 1 5 Explosion Protection 1...
Page 64: ...54733BA26 1 64 02 2017 5 Explosion Protection Volume 1...
Page 66: ...54733BA26 1 66 02 2017 5 Explosion Protection Volume 1...
Page 67: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 67 Volume 1 5 Explosion Protection 1...
Page 68: ...54733BA26 1 68 02 2017 5 Explosion Protection Volume 1...
Page 69: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 69 Volume 1 5 Explosion Protection 1...
Page 70: ...54733BA26 1 70 02 2017 5 Explosion Protection Volume 1...
Page 71: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 71 Volume 1 5 Explosion Protection 1...
Page 72: ...54733BA26 1 72 02 2017 5 Explosion Protection Volume 1...
Page 73: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 73 Volume 1 5 Explosion Protection 1...
Page 108: ...54733 10BA2L 1 108 02 2017 7 Functional Safety Volume 1 7 13 Functional Safety Certificate...
Page 110: ...54733 10BA2L 1 110 02 2017 7 Functional Safety Volume 1...
Page 112: ...54733 10BA2L 1 112 02 2017 8 Visual Inspection Volume 1...
Page 128: ...54733 10BA2L 1 128 02 2017 9 Radiation Protection Volume 1...
Page 147: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 1 147 Volume 1 1 Notes...
Page 149: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 2 Installing SENSseries...
Page 150: ...54733 10BA2L 2 150 02 2017 Volume 2...
Page 170: ...54733 10BA2L 2 170 02 2017 1 System Description Volume 2...
Page 240: ...54733 10BA2L 2 240 02 2017 4 Repair Maintenance and Upkeep Volume 2...
Page 288: ...54733 10BA2L 2 288 02 2017 5 Technical Information Volume 2...
Page 297: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 2 297 Volume 2 2 Notes...
Page 299: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 3 Operation with HART Communicator...
Page 300: ...54733 10BA2L 3 300 02 2017 Volume 3...
Page 390: ...54733 10BA2L 3 390 02 2017 3 Getting Started via the HART Communicator Volume 3...
Page 410: ...54733 10BA2L 3 410 02 2017 6 Functional Processes Volume 3...
Page 428: ...54733 10BA2L 3 428 02 2017 7 Explanations Volume 3...
Page 440: ...54733 10BA2L 3 440 02 2017 8 Error Handling Volume 3...
Page 444: ...54733 10BA2L 3 444 02 2017 9 Setup Protocol Volume 3 Path Parameters Unit Standard SETUP...
Page 445: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 3 445 Volume 3 3 Notes...
Page 449: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 1 449 Volume 1 Notes...