54733BA26
1 – 78
02.2017
5 Explosion Protection
Volume 1
Certificate:
70009819
Project:
70066628
Master Contract:
215040
Date Issued:
2016-03-07
DQD 507 Rev. 2012-05-22
Page 5
IIB
IIC
L
o
(mH)
C
o
(µF)
L
o
(mH)
C
o
(µF)
17.0
0.820
4.0
0.107
Or
Hart-current output (Sink Mode) (Terminal 17, 18)
Type of protection Intrinsic Safety Ex ia IIC Maximum Values:
U
i
= 30.0 V; I
i
= 152.0 mA; P
i
= 1.14 W; C
i
= 3.0 nF; L
i
= 20.0 µH
Note:
the HART current output (source mode or sink mode) of the current output module are safely electrically
isolated from earth and all other circuits.
Schedule of Limitations:
1)
The interface circuit RS485 serves exclusively for intercommunication of the probes and must not be
connected to an external RS485 circuit.
2)
In gas atmospheres for the choice of Group IIB or IIC for the intrinsically safe circuits, all intrinsically
safe circuits and the scintillation meter LB 480 series shall be fully operated in the selected Group IIB or
IIC.
APPLICABLE REQUIREMENTS
CSA C22.2 No. 60079-0:2011 - Explosive atmospheres. PART 0: Equipment - General requirements.
CSA C22.2 No. 60079-1:2011 - Explosive atmospheres. PART 1: Equipment protection of flameproof enclosures
“d” Second Edition.
CSA C22.2 No. 60079-7:2012
–
Explosive atmospheres. PART 7: Equipment protection by Increased Safety ‘e”
CSA C22.2 No. 60079-11:2014
–
Explosive atmospheres. PART 11: Equipment protection by Intrinsic Safety “i”
CSA C22.2 No. 60079-31:2012 -
–
Explosive atmospheres. PART 31: Equipment dust ignition protection by
enclosure “t”
CSA C22.2 No. 61010-1:12 - Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use - Part 1: General Requirements
UL 61010-1:2
nd
Edition - Safety Requirements for Electrical Equipment for Measurement, Control, and
Laboratory Use - Part 1: General Requirements
UL 60079-0:2013 - Explosive atmospheres. PART 0: Equipment - General requirements.
UL 60079-1 - 2009 - Explosive atmospher
es. PART 1: Equipment protection by flameproof enclosures “d”.
UL 60079-7:2008
–
Explosive atmospheres. PART 7: Equipment protection by Increased Safety ‘e”
UL 60079-11:2013
–
Explosive atmospheres. PART 11: Equipment protection by Intrinsic Safety “i”
UL 60079-31:2015 -
–
Explosive atmospheres. PART 31: Equipment dust ignition protection by enclosure “t”
MARKINGS
The manufacturer is required to apply the following markings:
Summary of Contents for LB 480
Page 2: ......
Page 11: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 1 Safety Manual...
Page 12: ...54733 10BA2L 1 12 02 2017 Volume 1...
Page 22: ...54733 10BA2L 1 22 02 2017 4 Transport and Assembly Volume 1...
Page 56: ...54733BA26 1 56 02 2017 5 Explosion Protection Volume 1...
Page 57: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 57 Volume 1 5 Explosion Protection 1...
Page 58: ...54733BA26 1 58 02 2017 5 Explosion Protection Volume 1...
Page 59: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 59 Volume 1 5 Explosion Protection 1...
Page 60: ...54733BA26 1 60 02 2017 5 Explosion Protection Volume 1...
Page 61: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 61 Volume 1 5 Explosion Protection 1...
Page 62: ...54733BA26 1 62 02 2017 5 Explosion Protection Volume 1...
Page 63: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 63 Volume 1 5 Explosion Protection 1...
Page 64: ...54733BA26 1 64 02 2017 5 Explosion Protection Volume 1...
Page 66: ...54733BA26 1 66 02 2017 5 Explosion Protection Volume 1...
Page 67: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 67 Volume 1 5 Explosion Protection 1...
Page 68: ...54733BA26 1 68 02 2017 5 Explosion Protection Volume 1...
Page 69: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 69 Volume 1 5 Explosion Protection 1...
Page 70: ...54733BA26 1 70 02 2017 5 Explosion Protection Volume 1...
Page 71: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 71 Volume 1 5 Explosion Protection 1...
Page 72: ...54733BA26 1 72 02 2017 5 Explosion Protection Volume 1...
Page 73: ...SENSseries LB 480 BERTHOLD TECHNOLOGIES GmbH Co KG 1 73 Volume 1 5 Explosion Protection 1...
Page 108: ...54733 10BA2L 1 108 02 2017 7 Functional Safety Volume 1 7 13 Functional Safety Certificate...
Page 110: ...54733 10BA2L 1 110 02 2017 7 Functional Safety Volume 1...
Page 112: ...54733 10BA2L 1 112 02 2017 8 Visual Inspection Volume 1...
Page 128: ...54733 10BA2L 1 128 02 2017 9 Radiation Protection Volume 1...
Page 147: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 1 147 Volume 1 1 Notes...
Page 149: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 2 Installing SENSseries...
Page 150: ...54733 10BA2L 2 150 02 2017 Volume 2...
Page 170: ...54733 10BA2L 2 170 02 2017 1 System Description Volume 2...
Page 240: ...54733 10BA2L 2 240 02 2017 4 Repair Maintenance and Upkeep Volume 2...
Page 288: ...54733 10BA2L 2 288 02 2017 5 Technical Information Volume 2...
Page 297: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 2 297 Volume 2 2 Notes...
Page 299: ...1 g l Time 0 00 10 30 11 00 11 3012 00 12 30 Volume 3 Operation with HART Communicator...
Page 300: ...54733 10BA2L 3 300 02 2017 Volume 3...
Page 390: ...54733 10BA2L 3 390 02 2017 3 Getting Started via the HART Communicator Volume 3...
Page 410: ...54733 10BA2L 3 410 02 2017 6 Functional Processes Volume 3...
Page 428: ...54733 10BA2L 3 428 02 2017 7 Explanations Volume 3...
Page 440: ...54733 10BA2L 3 440 02 2017 8 Error Handling Volume 3...
Page 444: ...54733 10BA2L 3 444 02 2017 9 Setup Protocol Volume 3 Path Parameters Unit Standard SETUP...
Page 445: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 3 445 Volume 3 3 Notes...
Page 449: ...SENSseries LB 480 Level measurement BERTHOLD TECHNOLOGIES GmbH Co KG 1 449 Volume 1 Notes...