PN B10093AE
vi
Safety Notice
Electrical Safety
Be sure to use the anchoring system to secure the centrifuge in place. The anchoring system is
designed to reduce the possibility of injury or damage that could result from instrument movement
in the event of a major rotor mishap.
Do not replace any centrifuge components with parts not specified for use on this instrument.
Electrical Safety
To reduce the risk of electrical shock, this instrument uses a three or five-wire electrical cord and
plug to connect this equipment to earth-ground. To preserve this safety feature:
•
Make sure that the matching wall outlet receptacle is properly wired and earth-grounded.
Check that the line voltage agrees with the voltage listed on the name-rating plate affixed to
the centrifuge.
•
Never use a three-to-two wire plug adapter.
•
Never use a two-wire extension cord or a two-wire non-grounding type of multiple-outlet
receptacle strip.
Do not place containers holding liquid on or near the chamber door. If they spill, liquid may get into
the instrument and damage electrical or mechanical components.
Safety Against Risk of Fire
This instrument is not designed for use with materials capable of developing flammable or
explosive vapors. Do not centrifuge such materials (for example, chloroform or ethyl alcohol) in
this instrument nor handle or store them near the centrifuge.
Mechanical Safety
For safe operation of the equipment, observe the following:
•
Use only the Beckman Coulter rotors and accessories designed for use in this instrument.
•
Do not exceed the maximum rated speed of the rotor in use.
•
NEVER attempt to slow or stop a rotor by hand.
•
Do not lift or move the centrifuge while a rotor is spinning.
•
NEVER attempt to override the door interlock system while the rotor is spinning.
•
Do not lean on the centrifuge or place items on it while it is operating.
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......