Troubleshooting
Accessing the Rotor in Case of Power Failure
3
a.
Use a back-and-forth motion as you pull; this takes quite a bit of force.
The vacuum will be completely vented several seconds after the hose is detached.
NOTE
Several tubing lines are tied to the red hose. As you grasp the hose, place your hand over these
lines. Make sure that they stay connected as you pull.
CAUTION
Do not twist the hose to the left (counterclockwise) as you pull the hose, or you
will unscrew the hose fitting.
Figure 3.3
Venting the Chamber Vacuum
8
Replace the hose by pushing it down over the fitting as far as it can go.
1.
Vacuum Pump
2.
Grasp red hose and pull up (Do
not
disconnect
the tubing lines tied to the red hose.)
3.
Red hose removed from fitting
4.
Replace red hose
1
2
A
3
B
4
C