PN B10093AE
2-10
Operation
Manual Operation
3
Press the next function key or press
ENTER
, then
START
.
•
After you start the run, the
HOLD
indicator illuminates in the
TIME
display.
•
The elapsed run time is displayed.
•
The run continues until
STOP
is pressed.
Entering Run Temperature
Run temperature can be set from –10 to +40°C. If no value is entered, the last-entered temperature
is used.
1
Press
TEMP.
The
Temp:
field flashes.
2
Use the keypad to enter the required temperature.
Entered temperature appears in the
Temp:
field.
3
Press the
TEMP
key again to move the cursor to the
Max Temp:
field.
•
Entered temperature plus 4°C appears in the
Max Temp:
field, and
YES
and
NO
soft keys
appear.
•
The default maximum temperature is the set temperature plus 4°C.
•
You can specify a different maximum temperature, from +2°C above the set
temperature to 44°C.
•
You can also specify whether or not you want the system to shut down if the maximum
temperature is reached.
•
The centrifuge will always shut down if the system temperature reaches 50°C.
4
Use the keypad to enter a different maximum temperature, or accept the default.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Rotor : JA-30.50
Speed : 26 000 RPM = 81 800 x G
Time : 00:00
Temp: - - Max Temp : - - °C
Accel : MAX Decel : MAX
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......