PN B10093AE
4-4
Care and Maintenance
Circuit Breaker and Fuses
2
To remove the air filter, hold the side edges and lift the filter straight up until the bottom edge
is above the centrifuge bottom retaining strip.
a.
Pull the filter out, bottom edge first, and discard it.
3
Install a new filter (885218).
a.
The filter has a directional arrow on one of its edges; install the filter with this arrow
pointing toward the centrifuge.
b.
Holding the filter by the side edges, insert the top half behind the frame edge and lift up
until the bottom half clears the retaining strip.
c.
Then set the bottom edge down.
Circuit Breaker and Fuses
There are no user-replaceable fuses in the Avanti J-26S XPI centrifuge.
If the centrifuge circuit breaker trips for any reason, the power switch will move to the OFF (
O
)
position.
1
Reset the circuit breaker by turning the power switch back to the ON (
I
) position.
a.
If it trips again immediately, do not reset it.
Contact Beckman Coulter Field Service.
*
1.
Frame Edge
2.
Air Filter
3.
Retaining Strip
4.
Frame Edge
4
3
2
1
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......