Description
Controls and Indicators
Setup Screen
The setup screen (
) contains editable fields for parameter entry. The six function keys to
the left of the screen are for selecting parameter fields. The six soft keys to the right of the screen
are used to select options based on which function key is pressed. Guidelines direct the eye from
each function and soft key to the display line controlled by that key. A message line at the bottom
of the screen shows help and diagnostic messages during setup and operation.
Figure 1.4
The Setup Screen
When the centrifuge power is turned on, the setup screen displays the parameters of the last run
performed.
Function Keys
Pressing
ENTER
or another function key saves parameter entries.
TIME
•
In time mode, indicates the remaining run time in hours and minutes. The
TIME
display begins counting down when
START
is pressed and continues counting down
to 0, when deceleration begins.
•
In hold mode, indicates the time elapsed since
START
was pressed. After 99 hours and
59 minutes the timer will reset to 0 and continue counting elapsed time.
•
In
2
t
mode, indicates the user-entered
2
t
(accumulated centrifugal force on the
sample) and the calculated time remaining until the
2
t
setting is reached.
TEMP°
C
Indicates rotor temperature in degrees C. The temperature is displayed to 1 degree and
is within
2 degrees of the set temperature (after temperature equilibrium is reached).
1.
Function Keys
2.
Message Line
3.
Soft Keys
1
2
3
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......