PN B10093AE
1-9
Description
Specifications
1
Control Features
Operational Features
Specifications
Description
Speed
•
Setting range
:
—
100 to 26,000 RPM
or
equivalent RCF (in 100
g
increments)
•
Elutriation speed setting range
:
—
0 to 5000 RPM (in 10-RPM or 10
g increments)
•
Speed display
:
—
from 0 to 10,000 RPM, display indicates actual rotor speed
10 RPM; from 10,000 to
26,000 RPM, indicates actual rotor speed
0.1%,
or
equivalent RCF
Speed Control
•
Low speed (2000 RPM) control = ±10 RPM
•
High speed (Max/Rotor) control = ±0.1% RPM
Time
•
Setting range
:
—
1 min to 99 hr 59 min,
2
t
, or continuous (hold)
•
Actual display
:
—
Indicates time remaining (timed run),
2
t
, or elapsed time (hold run)
•
2
t
setting range
:
—
up to 9.99
10
14
radians squared per second
•
2
t
actual display
:
—
indicates accumulated centrifugal effect to 3 significant digits (in exponential notation)
Temperature
•
Setting range:
—
–10 to +40°C (in 1°C increments)
•
Accuracy
:
—
rotor temperature controlled to within
2°C of set temperature (after equilibration)
a
•
Ambient temperature range
:
—
16 to 38°C (60 to 100°F)
b
•
Cooling fluid
:
—
refrigerant R452A (HFO)
Acceleration
maximum, slow, or timed (1 to 10 min from 0 to 500 RPM)
Deceleration
maximum, slow, timed (1 to 10 min from 500 to 0 RPM), or off
a. During transient conditions, such as acceleration and deceleration, rotor temperature may be outside this range. Refer to applicable rotor
manuals for specific rotor operating range information.
b. To reach temperatures above ambient, the centrifuge is dependent on the frictional heat generated inside the chamber during operation.
At low run speeds or low ambient temperatures, the centrifuge may not be able to achieve some higher temperatures.
Specifications
Description
Door
6.1-cm (2.4-in.) thick plastic resin with steel plate
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......