Troubleshooting
Accessing the Rotor in Case of Power Failure
4
Lift the front panel up and off the centrifuge; set it aside.
•
You will see an inner front panel that extends about halfway down from the centrifuge top
(see
•
Below this inner panel is a black manual door release interlock lever (see
Figure 3.2
Manual Door Release
5
Pull the interlock lever out and to the left (at about a 45-degree angle), and while holding it out,
step on the foot pedal.
Depending on the level of vacuum in the chamber at the time of the power failure, the door may
or may not open.
a.
If it opens, go to step 6.
b.
If it does not open, you will need to manually vent the vacuum. Go to step 7.
6
After the door opens, first release the foot pedal and then release the interlock lever.
Sample can be removed.
a.
Go to step 10.
7
To vent the chamber vacuum, grasp the red rubber hose and pull it up until it comes off of the
pump fitting (see
).
1.
Inner Front Panel
2.
Retaining Strip
3.
Manual Door Release
Interlock Lever
1
2
3
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......