PN B10093AE
3-3
Troubleshooting
User Messages
3
D1–5
/
Drive error, call
Beckman Coulter Field
Service
Drive system problem (for
example, over current,
over/under voltage,
components over
temperature)
Run stops, usually with
no brake. Door may not
unlock for up to an hour.
Call Beckman Coulter Field Service.
a
Before trying to open the door, listen
carefully and make sure that no sound
is coming from the chamber (indicating
a spinning rotor). Follow the directions
Accessing the Rotor in Case of
F1,
F4
,
F5
/
FRS
b
failure, refer to
manual
Required FRS level
not reached
Run shuts down with
maximum brake
•
Check and clean door sealing area
and door gasket.
•
Wipe any ice and excess moisture
from chamber.
•
Check and clean door sealing area
and door gasket.
F2
/
FRS failure, refer to
manual
FRS level too high
Run continues
Press
CE
. If F2 message reappears,
perform same actions as for Fl error.
F3
/
FRS failure, refer to
manual
FRS level too high
Run shuts down with
maximum brake
Call Beckman Coulter Field Service.
a
S1
/
System error
There is a problem with the
system control software,
EPROM, or RAM
System shuts down
Call Beckman Coulter Field Service.
a
H4
/
Speed error
•
a run is in progress (at
speed or accelerating)
•
the operator changes
the speed to 0 RPM by
pressing
SPEED,
0
,
ENTER
rather than by
STOP
key
•
the operator then tries
to increase the speed
by pressing
SPEED
,
entering a new speed,
and pressing
ENTER
.
The centrifuge does not
accelerate, and the H4
diagnostic occurs.
(Note that during a run,
the speed can be
changed from any speed
other than 0 by pressing
SPEED
, entering the new
speed, and pressing
ENTER
.)
Press
STOP
(green light on
START
key
will turn off), then press
CE
to clear the
diagnostic. Set the new run speed and
then press
ENTER
,
START
to begin a
new run.
Table 3.1
Diagnostic Message Chart
(Continued)
Diagnostic Message Problem
Result
Recommendation
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......