Troubleshooting
Introduction
This section lists possible malfunctions, along with probable causes and corrective actions. Maintenance
procedures are given in
CHAPTER 4, Care and Maintenance
. For any problems not covered here, contact
Beckman Coulter Field Service (1-800-742-2345 in the United States; outside the U.S. contact your local
Beckman office or visit us at www.beckman.com) for assistance.
NOTE
It is your responsibility to decontaminate the instrument, as well as any rotors and/or accessories,
before requesting service by Beckman Coulter Field Service.
User Messages
User messages appear on the setup screen to communicate information about the instrument or to
alert you to abnormal conditions that need attention.
•
Help and informational messages caused by incorrect input or certain operating conditions can
be cleared by pressing
CE
and following the instructions in the message.
•
Diagnostic messages that result from abnormal operating conditions or equipment
malfunction require troubleshooting. Refer to
to determine the nature of the
condition and any recommended actions. If a problem persists after you have performed the
recommended action, call your Beckman Coulter Field Service representative (in the United
States, call 1-800-742-2345; outside the U.S., call your local Beckman Coulter office). To help the
field service representative diagnose and correct the problem, try to gather as much
information about the situation as you can, including:
—
the diagnostic number and message,
—
the operating situation when the diagnostic condition occurred (such as rotor in use,
speed, or load type), and
—
any unusual environmental and/or operating conditions (such as ambient temperature or
voltage fluctuations).
NOTE
is a user guide, and is not a comprehensive checklist.
Summary of Contents for Avanti J-26S XPI
Page 4: ...PN B10093AE iv Revision History ...
Page 12: ...xii Contents Avanti J Series Centrifuge Warranty Related Documents ...
Page 14: ...xiv Illustrations ...
Page 16: ...xvi Tables ...
Page 20: ...PN B10093AE xx Introduction Multi Compliance Label ...
Page 36: ...PN B10093AE 1 16 Description Available Rotors ...
Page 66: ...PN B10093AE 3 10 Troubleshooting JCF Z Rotor Identification ...
Page 82: ...PN B10093AE A 10 Preinstallation Requirements Using J2 Series Rotors in the Avanti J 26S XPI ...
Page 86: ...PN B10093AE Warranty 2 Beckman Coulter Inc Avanti J Series Centrifuge Warranty ...
Page 87: ......