Configuring and operating the system
Syscompact 2000 M pro
48 / 98
822-175-2
10.6 Displaying system information
Prerequisite
The system is switched on.
Further information: Chapter
Switching on the system
(on page 40)
Procedure
1. Press the
MENU
key on the display.
2. Select the
INSTRUMENT SETTING
menu item and press the
ENTER
key.
3. Select the
INFO
menu item and press the
ENTER
key.
The firmware version, the version of Card A and the date of the last calibration are
displayed.
10.7 Switching the high voltage on and off
Switching on the high voltage
1. Release the high voltage: Press the
key for at least one second.
The system status changes to the
Ready to switch on
operating state. The red indicator
light flashes for approx. 6 seconds.
2. Switch on the high voltage by pressing the
key within 6 seconds (while the red
indicator light flashes).
3. The system status changes to the
In operation
operating state.
If you do not press the
key within this period of time, the system will be switched back
to the
Ready for operation
operating state.
Switching off the high voltage
1. Press the
key.
DANGER
Dangerous voltage on the test object and other live machine parts.
Danger to life,
risk of injury from high electric voltage.
2. Before touching test object, discharge, earth and short-circuit the same: at the connection
point and at the far end.
You may touch the plant parts that were under voltage only if they are visibly earthed and
short-circuited.
Summary of Contents for Sys compact 2000 M pro
Page 97: ......