System manual DICOM Theater
73
8 For all the devices in the configuration, define (if necessary) and
select a Preset.
9 After selecting the Preset, MediCal Pro starts consistency
calibration automatically.
Please refer to the MediCal Pro User Guide for more information.
OPTICAL SENSOR CONNECTION
The optical sensor can be attached to the screen by means of Velcro,
attached to the rear of the sensor.
The sensor must be connected to the PC through USB.
Note: When you plug in the optical sensor to the PC USB bus for the
first time, you will be asked to install a driver. You can find the driver on
the Dicom Theater application CD-ROM, in the folder Drivers >
Sequel.
To connect the sensor:
1 Before attaching the sensor, be sure to set the sensor patch to the
desired location (see above). You can always change the position
of the patch by means of MediCal Pro.
2 Peel off the protective sheet of the Velcro part that will be attached
to the screen.
Landscape projection
Portrait projection
3 Attach the sensor to the screen, in the center of the optical sensor
patch.
Summary of Contents for DICOM THEATER
Page 5: ...System manual DICOM Theater 6 ...
Page 7: ...System manual DICOM Theater 8 INTRODUCTION ...
Page 8: ...System manual DICOM Theater 9 This page intentionally left blank ...
Page 17: ...System manual DICOM Theater 18 IMAGING BOARD INSTALLATION ...
Page 18: ...System manual DICOM Theater 19 This page intentionally left blank ...
Page 25: ...System manual DICOM Theater 26 BARCOMED SOFTWARE DRIVER INSTALLATION ...
Page 26: ...System manual DICOM Theater 27 This page intentionally left blank ...
Page 56: ...System manual DICOM Theater 57 This page intentionally left blank ...
Page 62: ...System manual DICOM Theater 63 Figure 28 Windows NT 4 0 Screen Saver Tab ...
Page 69: ...System manual DICOM Theater 70 MEDICAL AND DICOM THEATER SOFTWARE INSTALLATION AND USAGE ...
Page 70: ...System manual DICOM Theater 71 This page intentionally left blank ...