System manual DICOM Theater
13
Installation examples
Single-head installation
Proceed as follows:
1 Connect imaging board Head A to the display VGA video input by
means of the D-Sub15-to-D-Sub15 cable.
2 Connect imaging board Head B (marked “Projector out”) to the
projector BNC input by means of the D-Sub15-to-5 BNC cable (15
m / 50 ft).
3 Connect a free PC COM port to the projector RS-232 connector by
means of the D-Sub9-to-D-Sub9 cable (15 m / 50 ft).
4 Connect a free PC USB connector to the display USB In connector
by means of a USB-to-USB cable.
5 Connect the optical sensor to another free PC USB bus. If the cable
is not long enough, you have to use a USB booster/extender.
Installation of the sensor is described more in detail in the section
“MediCal and Dicom Theater software installation and usage”.
Video out
BNC in
VGA in
USB out
USB in
COM port
RS-232 in
Optical sensor
Head B
Head A
Summary of Contents for DICOM THEATER
Page 5: ...System manual DICOM Theater 6 ...
Page 7: ...System manual DICOM Theater 8 INTRODUCTION ...
Page 8: ...System manual DICOM Theater 9 This page intentionally left blank ...
Page 17: ...System manual DICOM Theater 18 IMAGING BOARD INSTALLATION ...
Page 18: ...System manual DICOM Theater 19 This page intentionally left blank ...
Page 25: ...System manual DICOM Theater 26 BARCOMED SOFTWARE DRIVER INSTALLATION ...
Page 26: ...System manual DICOM Theater 27 This page intentionally left blank ...
Page 56: ...System manual DICOM Theater 57 This page intentionally left blank ...
Page 62: ...System manual DICOM Theater 63 Figure 28 Windows NT 4 0 Screen Saver Tab ...
Page 69: ...System manual DICOM Theater 70 MEDICAL AND DICOM THEATER SOFTWARE INSTALLATION AND USAGE ...
Page 70: ...System manual DICOM Theater 71 This page intentionally left blank ...