TBN581 (ORP/pION) TRANSMITTER
CALIBRATION
I - 6
WBPEEUI110501A0
This value is recalculated during each calibration.
This value shows the offset change over time for a particu-
lar sensor. Value resets to 0.00 pH if installing a new sen-
sor.
Calibration complete.
Key
Display
Comments
Select
BENCH CAL.
SAME SENSOR:CALCULATED SENSOR OFFSET
,
CAL-
CULATED EFFICIENCY
, and
CUMULATIVE OFFSET
CHANGE
are recalculated.
NEW SENSOR: CALCULATED SENSOR OFFSET, CAL-
CULATED EFFICIENCY
are calculated and
CUMULA-
TIVE OFFSET CHANGE
is set to 0.00 mV.
NOTE:
mV SOURCE
selection is covered in
Low default value: -1000 mV.
High default value: +1000 mV.
Enter low mV value and press
. Enter high mV
value.
Key
Display
Comments
ENTER
CALCULATED
SENSOR OFFSET
nnn.nn pH
HIT ENTER
ENTER
CUMULATIVE
OFFSET CHANGE
n.nn pH
HIT ENTER
ENTER
CONFIG ID
READY
CALI-
BRATE
OUTPUT WILL BE
AFFECTED
PROCEED?
NO
→
YES
ENTER
CALIBRATION
PROCESS CAL
→
BENCH CAL
D-TO-A ADJUST
ENTER
INPUT SOURCE:
→
SAME SENSOR
NEW SENSOR
mV SOURCE
ENTER
ENTER LO mV CAL
nn.nn mV
ENTER HI mV CAL
nn.nn mV
ENTER