TBN581 (ORP/pION) TRANSMITTER
CALIBRATION
WBPEEUI110501A0
I - 7
Bench Calibration (mV)
This procedure applies to TBN581 Transmitters configured for either
ANA-
LOG
or
DIGITAL
. Any difference between the two are noted in the
Com-
ments
column.
Displays low mV calibration value, and the uncalibrated
mV values determined from the input source.
Displays high mV calibration value, and the uncalibrated
mV values determined from the input source.
Value from last calibration. If new sensor, efficiency will
be 98%.
This value is recalculated during each calibration.
This value shows the offset change over time for a particu-
lar sensor. Value resets to 0.00 mV if installing a new sen-
sor.
Calibration complete.
Key
Display
Comments
ENTER
APPLY nn.nn mV
nn.nn
HIT ENTER
WHEN STABLE
ENTER
APPLY nn.nn mV
nn.nn
HIT ENTER
WHEN STABLE
ENTER
CALCULATED
EFFICIENCY
nnn.n %
HIT ENTER
ENTER
CALCULATED
SENSOR OFFSET
nnn pH
HIT ENTER
ENTER
CUMULATIVE
OFFSET CHANGE
nnn pH
HIT ENTER
ENTER
CONFIG ID TAG
READY
Key
Display
Comments
CALI-
BRATE
OUTPUT WILL BE
AFFECTED
PROCEED?
NO
→
YES