TROUBLESHOOTING
ERROR MESSAGES
WBPEEUI110501A0
5 - 3
FIELD DEVICE
CAN'T EXECUTE
COMMAND:
FBS ON-LINE
All
STT03E terminal cannot
change any transmitter
parameters while the field
bus is active.
Transmitter must be brought off-line. This
can be done from the MFC/MFP by tuning
the function code.
NOTE:
To bring transmitter off-line, the retainer can be
unplugged from the rack. Be aware that by doing this, com-
munication is lost for all other transmitters on the bus.
When the retainer is unplugged, there is a 2-minute time out
period until the STT03E terminal will be allowed to change
parameters. After the 2-minute period, the transmitter al-
lows the STT03E terminal access.
FIELD DEVICE
CAN'T EXECUTE
COMMAND:
HARDWARE PROBLEM
BCN
EQN
TBN
Detected error in
hardware.
Press
to determine error and use
this table for appropriate corrective action. If
status is not more explicit, check electronics
assembly of field device.
AVS
Unspecified hardware
failure
Service unit.
FIELD DEVICE
CAN'T EXECUTE
COMMAND:
INVALID COMMAND
All
Incorrect syntax.
Wrong device type selected in the
configuration. Refer to the appropriate device
configuration appendix for details about
device type.
Excessive line noise.
Check line noise. Use an oscilloscope to
determine if line noise is excessive.
STT03E terminal failure.
Verify STT03E terminal is operational. Verify
by using another STT03E terminal that is
known to be operational.
FIELD DEVICE
CAN'T EXECUTE
COMMAND:
LOCKOUT ENGAGED
All
Unable to configure or cali-
brate. Hardware lock active.
Refer to the device product instruction for
details about the configuration lockout
jumper.
FIELD DEVICE
CAN'T EXECUTE:
DATA
OUT OF RANGE
All
Transmitter sent data that is
out of acceptable range.
Press
to verify the parameters of the
configuration and consult the appropriate
device instruction for correct parameters.
FIELD DEVICE
CONFIGURATION
DOES NOT MATCH
STT03'S
All
Configuration of the trans-
mitter does not exactly
match the corresponding
configuration in the STT03E
terminal.
Rerange was executed without updating the
STT03E terminal configuration.
Perform
GET CONFIG
. Refer to
NOTE:
The AVS will change the installed option flags if
they are incorrect in the transmitted configuration. This er-
ror will appear if the configuration is not reloaded by the
STT03.
FIELD DEVICE
MAIN
ELECTRONICS
EEPROM FAILURE
BCN
EQN
TBN
Memory problem.
Reconfigure and recalibrate the transmitter.
Refer to the appropriate configuration and
calibration sections in this instruction for the
transmitter type.
EQS
PTS
Memory problem.
Remove power from the field device and
reapply. If problem persists replace the ampli-
fier assembly. Refer to repair/replacement
section of the field device product instruction.
Table 5-1. STT03E Error Messages
(continued)
Message
AffectedD
evice
Probable Cause
Corrective Action
STATUS
VIEW