TBN580 TRANSMITTER
CREATE/MODIFY CONFIGURATION
WBPEEUI110501A0
H - 3
Select
pH TBN
.
For this example, select
STANDARD
.
STANDARD
: thermo-isopotential 7.0 pH;
slope 59.15 mV at 25
°
C.
ANTIMONY
: isopotential 1.5 pH; slope 51 mV
at 25
°
C.
NOTE:
If after selecting
ANTIMONY,
the STT03E termi-
nal branches to a
TEMPERATURE DISPLAY UNITS
selection screen, choose the desired temperature unit.
Select
3.01 BALCO.
When
NONE
is selected, the STT03E terminal branches to
a
TEMPERATURE DISPLAY UNITS
selection screen.
Choose the desired temperature unit.
For this example select
AUTO SOLUTION.
NOTE:
Other temperature compensation selections
branch to a
TEMPERATURE DISPLAY UNITS
selection
screen.
This screen is unique to the
AUTO SOLUTION
selection
of the
TEMP COMP TYPE
. Other
selections go to the temperature display units.
If
YES
is selected, enter 5 input and output points. Values
must be between 0 and 100%.
Key
Display
Comments
ENTER
TRANSMITTER TYPE
cond TBN
→
pH TBN
ORP/pION
ENTER
pH SENSOR INPUT
→
STANDARD
ANTIMONY
ENTER
TEMP COMP INPUT
→
3.01 K BALCO
NONE
ENTER
TEMP COMP TYPE
MANUAL
AUTO NERNSTIAN
→
AUTO SOLUTION
AUTO PURE H2O
ENTER
ENTER SOLUTION
CO-EFFICIENT
±
n.nnn pH per
10
°
C (from 25
°
C)
ENTER
TEMPERATURE
DISPLAY UNITS
→ °
C
°
F
ENTER
OUTPUT FUNC GEN
→
NO
YES