TBN580 TRANSMITTER
CALIBRATION
H - 10
WBPEEUI110501A0
D-to-A Adjust
The
D-TO-A ADJUST
selection is only present when you are in the
ANA-
LOG
communication mode, specified in the configuration procedure. This
selection allows you to adjust the four to 20 milliamphere output of the field
device.
Select
BENCH CAL TEMP.
If in
TEMP COMP TYPE
,
MANUAL
or
NONE
is selected
during configuration, this screen will not appear, other-
wise the next screen,
CHANGE TEMP OF PROCESS?
appears. Default is 25
°
C.
The value displayed is the temperature of the process
(interior of sensor).
NOTE:
If a new temperature value is entered, this change
will affect the calibration of temperature.
Calibration complete.
Key
Display
Comments
ENTER
CALIBRATION
PROCESS CAL
BENCH CAL pH
→
BENCH CAL TEMP
ENTER
ENTER FIXED
PROCESS TEMP
nnn.n
°
UNITS
ENTER
CHANGE TEMP
OF PROCESS?
nnn.n
°
UNITS
ENTER
CONFIG. ID TAG
READY
Key
Display
Comments
CALI-
BRATE
OUTPUT WILL BE
AFFECTED
PROCEED?
NO
→
YES