TBN580 TRANSMITTER
CALIBRATION
WBPEEUI110501A0
H - 7
Bench Calibration pH
The following procedure steps through the bench calibration of the pH level.
This value is recalculated during each calibration.
This value shows the offset change over time for a particu-
lar sensor. Value resets to 0.00 pH if installing a new sen-
sor.
Calibration complete.
Key
Display
Comments
ENTER
CURRENT READING
nn.nn pH
ENTER DESIRED
nn.nn pH
ENTER
CALCULATED
EFFICIENCY
nnn.nn%
HIT ENTER
ENTER
CALCULATED
SENSOR OFFSET
nnn.nn pH
HIT ENTER
ENTER
CUMULATIVE
OFFSET CHANGE
n.nn pH
HIT ENTER
ENTER
CONFIG. ID TAG
READY
Key
Display
Comments
CALI-
BRATE
OUTPUT WILL BE
AFFECTED
PROCEED?
NO
→
YES