EQS TEMPERATURE TRANSMITTER
CALIBRATION
D - 6
WBPEEUI110501A0
Key
Display
Comments
Select
YES.
Select
BENCH CALIB
with the down arrow key.
Enter the low and high calibration voltages
.
Enter the low and high calibration resistances.
Apply 0.00 mV to the transmitter.
NOTE:
Refer to the EQS instruction for detailed wiring
diagrams of this procedure.
Apply 100 mV to the transmitter.
NOTE:
Refer to the EQS instruction for detailed wiring
diagrams of this procedure.
Apply 0
Ω
to the transmitter.
NOTE:
Refer to the EQS instruction for detailed wiring
diagrams of this procedure.
Apply 250
Ω
to the transmitter.
NOTE:
Refer to the EQS instruction for detailed wiring
diagrams of this procedure.
CALI-
BRATE
OUTPUT WILL BE
AFFECTED!
PROCEED?
NO
→
YES
ENTER
CALIBRATION
D-TO-A ADJUST
→
BENCH CALIB.
ENTER
LOW CALIB
n.nnn mV
HIGH CALIB
n.nnn mV
ENTER
LOW CALIB RES
n.nn
Ω
HIGH CALIB RES
nnn.nn
Ω
ENTER
APPLY
0.00 mV
AT 3(+), 4(-)
THEN HIT ENTER
ENTER
APPLY
100.00 mV
AT 3(+), 4(-)
THEN HIT ENTER
ENTER
APPLY 0 .00
Ω
SHORT
1-2, 2-3, 3-4
THEN HIT ENTER
ENTER
APPLY 250
Ω
SHORT 1-2, 3-4
Ω
AT 2-3
THEN HIT ENTER