Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are
compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used
for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Fail-
ure to do so can result in the following problems:

Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the applica-
tion/process solution only needs to be of pH >8 to cause visible surface changes.

Material damage such as corrosion, cracks, fracturing, premature aging or swelling.

Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause
corrosion.

Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at
www.a-k-i.org, link to "AKI-Brochures", "Red brochure".

3.3

Reusable products

No maximum number of applications and processing cycles has been specified for the product.

The service life of the product is limited by damage, normal wear, the type and duration of usage, as well as the
handling, storage and transportation of the product.

Careful visual and functional testing prior to next use is the best way to identify a malfunctioning product.

3.4

Preparations at the place of use

If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example.

Remove any visible surgical residues to the extent possible with a damp, lint-free cloth.

Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.

3.5

Preparing for cleaning

Do not disassemble any anchoring screws/bolts that permanently connect the various components. 

Disassemble the product prior to cleaning, see Disassembly.

3.6

Disassembly

Remove the movable measuring arm

a

 from the rod length measuring instrument.

3.7

Cleaning/Disinfection

3.7.1

Product-specific safety information on the reprocessing method

Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temper-
atures! 

Use cleaning agents and disinfectants in accordance with the manufacturer's instructions.

Observe specifications regarding concentration, temperature and exposure time.

Do not exceed the maximum allowable disinfection temperature of 95 °C.

Use suitable cleaning/disinfectants for wet disposal. To prevent foam from forming and reduced effectiveness of
the process chemicals: Prior to mechanical cleaning and disinfection, rinse the product thoroughly under running
water

3.7.2

Validated cleaning and disinfection procedure

3.8

Manual cleaning/disinfection

Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution.

After manual cleaning/disinfection, check visible surfaces visually for residues.

Repeat the cleaning/disinfection process if necessary.

3.8.1

Manual cleaning with immersion disinfection

D–W:

Drinking water

FD–W:

Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)

RT:

Room temperature

*Recommended: BBraun Stabimed fresh

Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.

Phase I

Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.

Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.

If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.

Mobilize non-rigid components, such as set screws, links, etc. during cleaning.

Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.

Phase II

Rinse/flush the product thoroughly (all accessible surfaces) under running water.

Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.

Drain any remaining water fully.

Phase III

Fully immerse the product in the disinfectant solution.

Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.

Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe.
Ensure that all accessible surfaces are moistened.

Phase IV

Rinse/flush the product thoroughly (all accessible surfaces).

Mobilize non-rigid components, such as set screws, joints, etc. during final rinse.

Rinse lumens with an appropriate disposable syringe at least five times.

Drain any remaining water fully.

Phase V

Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning
and disinfection procedure.

3.9

Machine cleaning/disinfection

Note

The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).

Note

The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.

3.9.1

Mechanical alkaline cleaning and thermal disinfection

Machine type: single-chamber cleaning/disinfection device without ultrasound

D–W:

Drinking water

FD–W:

Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)

*Recommended: BBraun Helimatic Cleaner alcaline

Check visible surfaces for residues after mechanical cleaning/disinfecting.

3.10

Mechanical cleaning/disinfection with manual pre-cleaning

Note

The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).

Note

The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.

3.10.1 Manual pre-cleaning with a brush

D–W:

Drinking water

RT:

Room temperature

*Recommended: BBraun Stabimed fresh

Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.

Phase I

Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.

Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.

If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.

Mobilize non-rigid components, such as set screws, links, etc. during cleaning.

Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.

Phase II

Rinse/flush the product thoroughly (all accessible surfaces) under running water.

Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.

Validated procedure

Specific requirements

Reference

Manual cleaning with immersion 
disinfection

FW728R to FW729R

FW731R to FW733R

FW737R to FW739R

Suitable cleaning brush

20 ml disposable syringe

Clean products having movable 
hinges in the open position or while 
moving the joints.

Drying phase: Use a lint-free cloth or 
medical compressed air 

Chapter Manual clean-
ing/disinfection and subsec-
tion:

Chapter Manual cleaning 
with immersion disinfec-
tion

Alkaline machine cleaning and 
thermal disinfection

FW729R

FW732R

FW733R

FW738R

Place product on a screen basket 
suitable for cleaning (make sure all 
areas will be reached by water jets).

Connect components with lumens 
and channels directly to the rinsing 
port of the injector carriage.

Open the links and joints before plac-
ing the product on the screen basket.

Chapter Machine clean-
ing/disinfection and sub-sec-
tions:

Chapter Mechanical alka-
line cleaning and thermal 
disinfection

Manual pre-cleaning with brush 
and subsequent mechanical alka-
line cleaning and thermal disinfec-
tion

FW728R

FW731R

FW737R

FW739R

Suitable cleaning brush

20 ml disposable syringe

Place the product on a tray that is 
suitable for cleaning (avoid rinsing 
blind spots).

Connect the lumens and channels of 
individual parts directly to the injec-
tor unit's special flushing connectors.

Open the links and joints before plac-
ing the product on the screen basket.

Chapter Mechanical clean-
ing/disinfection with manual 
pre-cleaning and subsection:

Chapter Manual pre-
cleaning with a brush

Chapter Mechanical alka-
line cleaning and thermal 
disinfecting

Phase

Step

T
[°C/°F]

t
[min]

Conc. 
[%]

Water 
quality

Chemical

I

Disinfecting clean-
ing

RT 
(cold)

>15

2

D–W

Aldehyde-free, phenol-free, and 
QUAT-free concentrate, pH ~ 9*

II

Intermediate rinse

RT 
(cold)

1

-

D–W

-

III

Disinfection

RT 
(cold)

5

2

D–W

Aldehyde-free, phenol-free, and 
QUAT-free concentrate, pH ~ 9*

IV

Final rinse

RT 
(cold)

1

-

FD-W

-

V

Drying

RT

-

-

-

-

Phase

Step

T
[°C/°F]

t
[min]

Water 
quality

Chemical/Note

I

Pre-rinse

<25/77

3

D–W

-

II

Cleaning

55/131

10

FD-W

Concentrate, alkaline: 

pH  =  13

<5 % anionic surfactant

0.5 % working solution

pH = 11*

III

Intermediate rinse

>10/50

1

FD-W

-

IV

Thermal disinfecting

90/194

5

FD-W

-

V

Drying

-

-

-

According to the program for cleaning 
and disinfection device

Phase

Step

T
[°C/°F]

t
[min]

Conc. 
[%]

Water 
quality

Chemical

I

Disinfec-
tant 

                      

cleaning

RT 
(cold)

>15

2

D–W

Aldehyde-free, phenol-free, and 
QUAT-free concentrate, pH ~ 9*

II

Rinsing

RT 
(cold)

1

-

D–W

-

Summary of Contents for Aesculap TA014587

Page 1: ...i Description technique Instruments S4 Long Tab es Instrucciones de manejo Descripci n t cnica Instrumental S4 Long Tab it Istruzioni per l uso Descrizione tecnica Strumenti S4 Long Tab pt Instru es d...

Page 2: ...acked worn or fractured components Always carry out a function test prior to each use of the product WARNING Danger of severe injuries from incorrectly positioning implants The depth markings and scal...

Page 3: ...omponents such as set screws joints etc during final rinse Rinse lumens with an appropriate disposable syringe at least five times Drain any remaining water fully Phase V Dry the product in the drying...

Page 4: ...length measuring instrument 3 13 Packaging Place the product in its holder or on a suitable tray Ensure that sharp edges are covered Package trays appropriately for the sterilization process e g in Ae...

Page 5: ...teril geliefert Fabrikneues Produkt nach Entfernung der Transportverpackung und vor der ersten Sterilisation reinigen 2 3 Anwendung GEFAHR Lebensgefahr durch Vorschieben des K Drahts in die Aorta Beim...

Page 6: ...auf achten dass alle zug nglichen Oberfl chen benetzt sind Produkt mit einer geeigneten Reinigungsb rste in der L sung so lange reinigen bis auf der Oberfl che keine R ckst nde mehr zu erkennen sind W...

Page 7: ...ay JG600 oder STERILIT I Tropf ler JG598 Zerlegbares Produkt zusammenbauen siehe Montage Produkt auf Funktion pr fen Alle beweglichen Teile z B Scharniere Schl sser Sperren Gleitteile usw auf vollst n...

Page 8: ...cas de p n tration du fil Kirschner dans l aorte Lors de l insertion d implants S4 ou d instruments S4 par le fil de guidage tenir compte des rep res sur le fil de guidage S assurer du positionnement...

Page 9: ...nte pendant au moins 15 min Veiller ce que toutes les surfaces accessibles soient humect es Nettoyer le produit dans la solution avec une brosse de nettoyage appropri e jusqu ce qu aucun r sidu ne soi...

Page 10: ...au proc d de st rilisation utilis p ex pour la st rilisation la vapeur spray d huile STERILIT I JG600 ou compte gouttes d huile STERILIT I JG598 Assembler les produits d montables voir Montage Contr...

Page 11: ...aorta Al introducir implantes S4 o instrumental S4 en la aguja gu a Preste atenci n a las marcas situadas en la gu a Compruebe que la posici n es la correcta por control radiogr fico Sujete la gu a co...

Page 12: ...ci n Fase I Sumergir todo el producto en la soluci n desinfectante con acci n limpiadora durante al menos 15 min Com probar que todas las superficies accesibles quedan humedecidas Limpiar el producto...

Page 13: ...IT I JG600 o lubricador por goteo STERILIT I JG598 Montar el producto por piezas ver Montaje Comprobar el funcionamiento del producto Comprobar que todas las piezas m viles p ej bisagras candados bloq...

Page 14: ...previa rimozione dell imballo da trasporto 2 3 Utilizzo PERICOLO Pericolo di morte a causa dell avanzamento del filo di Kirschner nell aorta Quando si introducono impianti S4 o strumenti S4 sul filo g...

Page 15: ...zione disinfettante ad azione detergente attiva per almeno 15 min accertandosi che tutte le superfici accessibili risultino inumidite Pulire il prodotto con uno spazzolino per pulizia idoneo nella sol...

Page 16: ...la ste rilizzazione a vapore olio spray STERILIT I JG600 oppure oliatore a goccia STERILIT I JG598 Assemblare il prodotto smontabile vedere Montaggio Verificare il funzionamento del prodotto Controlla...

Page 17: ...3 Utiliza o PERIGO Risco de morte devido ao avan o do fio K para a aorta Ao inserir implantes S4 ou instrumentos S4 pelo fio guia Ter em aten o as marca es no fio guia Assegurar a posi o correta com c...

Page 18: ...almente o produto na solu o desinfectante de limpeza activa durante pelo menos 15 min Para tal assegurar que todas as superf cies acess veis ficam molhadas Limpar o produto com uma escova adequada na...

Page 19: ...ar se o produto funciona corretamente Inspecionar todas as pe as articuladas por ex dobradi as fechos bloqueios pe as deslizantes etc quanto sua total liberdade de movimentos Verificar a compatibilida...

Page 20: ...schoon voordat u het voor het eerst ste riliseert 2 3 Gebruik GEVAAR Levensgevaar door doorschuiven van de K draad in de aorta Bij het inbrengen van S4 implantaten of S4 instrumenten over de voerdraad...

Page 21: ...ndergedom peld Zorg ervoor dat alle bereikbare oppervlakken bevochtigd worden Reinig het product met een geschikte reinigingsborstel in de oplossing totdat er op het oppervlak geen residuen meer te be...

Page 22: ...demonteerde product zie Montage Controleer de werking van het product Controleer alle bewegende delen bijv scharnieren sloten schuifdelen enz op volledige beweging Controleer de compatibiliteit met de...

Page 23: ...VARNING Risk f r personskador och eller felfunktion Belasta endast skruvarna axialt vid inf randet i benet Undvik sidokrafter under iskruvningen Anv nd inte ins ttningsinstrumentet f r l smutter FW733...

Page 24: ...och desinfektionsapparat som anv nds m ste underh llas och kontrolleras regelbundet 3 9 1 Maskinell alkalisk reng ring och termisk desinfektion Apparattyp Reng rings desinfektionsapparat med en kammar...

Page 25: ...ska service se Teknisk service 3 12 Montering Skjut den r rliga m tsk nkeln a p m tinstrumentet f r stavl ngd 3 13 F rpackning L gg produkten i avsedd beh llare eller i en l mplig tr dkorg Se till att...

Page 26: ...ng Tab 2 1 2 2 1 3 2 2 2 2 1 2 2 2 S4 Long Tab O82302 S4 Spinal System S4 Spinal System TA011187 S4 MIS TA014087 S4 TA012384 S4 TA013496 2 2 3 2 3 S4 S4 FW733R FW732R FW732R FW739R S4 Long Tab S4 Long...

Page 27: ...I Brochures Red brochure 3 3 3 4 6 3 5 3 6 a 3 7 3 7 1 95 C 3 7 2 3 8 3 8 1 BBraun Stabimed fresh I 15 1 5 II III 5 IV 5 V 3 9 FDA CE FW728R FW729R FW731R FW733R FW737R FW739R 20 FW729R FW732R FW733R...

Page 28: ...3 14 FW729R DIN EN 285 DIN EN ISO 17665 134 C 5 3 15 4 B Braun Aesculap Aesculap Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1601 Fax 49 7461 16 2887 E Mail ats ae...

Page 29: ...rolou VAROV N Nebezpe razu a nebo nespr vn ho fungov n rouby p i zav d n do kosti zat ujte pouze axi ln Zabra te p soben bo n ch sil b hem zav d n N stroj na vsazen areta n matice FW733R nepou vejte k...

Page 30: ...ov enou innost nap povolen FDA nebo ozna en CE na z klad normy DIN EN ISO 15883 Upozorn n Pou it iat c a desinfek n p stroj mus b t pravideln udr ovan a kontrolovan 3 9 1 Strojn alkalick i t n a tepe...

Page 31: ...te na p slu n skladovac m sto nebo do vhodn ho s tov ho ko e Zajist te ochranu ost n stroj S tov ko e zabalte p im en steriliza n mu postupu nap do steriln ch kontejner Aesculap Zajist te aby obal za...

Page 32: ...ew a ciwego dzia ania Przed ka dym u yciem nale y sprawdzi produkt pod k tem lu nych wygi tych p kni tych porysowa nych zu ytych lub od amanych element w Przed ka dym u yciem przeprowadza kontrol dzia...

Page 33: ...fekcyjnym za pomoc odpowiedniej strzykawki jednorazowej Faza II Produkt nale y ca kowicie przep uka i sp uka bie c wod wszystkie dost pne powierzchnie Podczas czyszczenia nale y porusza ruchomymi kom...

Page 34: ...dy sterylizacji np olejem w aerozolu STERILIT I JG600 w przypadku sterylizacji parowej albo za pomoc olejarki STERILIT I JG598 Zmontowa rozk adany produkt patrz Monta Sprawdzi dzia anie produktu Wszys...

Page 35: ...u pomocou r ntgenovej kontroly Skontrolujte d ku ty e pomocou r ntgenovej kontroly VAROVANIE Nebezpe enstvo zranenia a alebo chybnej funkcie Pri vlo en do kosti skrutky za a ujte len axi lne Zabr te p...

Page 36: ...postupy istenia a dezinfekcie 3 9 Strojov istenie dezinfekcia Ozn menie istiace a dezinfek n zariadenie mus ma preuk zate n innos napr osved enie FDA alebo ozna enie CE pod a DIN EN ISO 15883 Ozn meni...

Page 37: ...dlo te na pr slu n miesto skladovania alebo polo te na vhodn sie ov k Ubezpe te sa e ostria ktor s k dispoz cii s chr nen Sitkov ko e pre steriliza n proces spr vne zaba te napr do steriln ch n dob od...

Page 38: ...n nakliyat ambalaj n n kar lmas ndan sonra ve ilk sterilizasyondan nce temizleyin 2 3 Uygulama TEHL KE Kirschner telinin ah damar na itilmesi sonucu hayati tehlike riski vard r S4 implantlar n n veya...

Page 39: ...z r n gerekirse zeltide uygun bir temizlik f r as yard m yla y zeyde g r n r hi bir art k kalmayana kadar temizleyiniz Gerekti inde g zle g r lmeyen y zeyleri en az 1 dakika uygun bir temizlik f r as...

Page 40: ...yonunda STERILIT I ya spreyi JG600 ya da STERILIT I damlal k ya lay c JG598 Par alara ayr labilir r n birle tirin bkz Montaj r n n fonksiyon kontrol n yap n T m hareketli par alar rn mente eler kilitl...

Reviews: