10. Technical Service
DANGER
Danger to life of patients and users in case of malfunctions and/or
failure of protective measures!
►
Do not perform any servicing or maintenance work under any cir-
cumstances while the product is being used on a patient.
CAUTION
Modifications of medical devices may result in a loss of potential
guarantee/warranty claims and forfeiture of applicable licenses.
►
Do not modify the product.
►
Contact national B. Braun/Aesculap representative for service and
repair.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
Other service addresses can be obtained from the address indicated above.
11. Accessories/Spare parts
Sterile funnel or battery
removal aid cannot be
plugged in
Sterile funnel or battery
removal device not compati-
ble
Sterile funnel or battery removal aid
not plugged into battery compartment
Use suitable sterile funnel or battery removal aid
for GA331.
Sterile funnel or battery
removal device
deformed/defective
Sterile funnel or battery removal device
difficult or not attachable
Replace sterile funnel or battery removal aid.
Battery compartment on
product deformed/defective
Difficult or impossible to plug in the
sterile funnel or battery removal aid
Have the product repaired by the manufacturer.
Speed control knob can-
not be actuated
Product in locked OFF position Safety catch is in OFF position
Turn the safety catch into the ON position.
Speed control knob
jammed/defective
Speed control knob not working
Have the manufacturer repair the product.
Saw blade not moving
Loud running noise from
motor
Broken gear unit
Have repaired by manufacturer
Insufficient cutting per-
formance of the saw
blade
Saw blade teeth worn
Blunt saw blade
Replace saw blade
Battery too weak
Insufficient motor power/speed
Charge battery
Battery worn/defective
Replace battery
Saw blade overheated
Chips not removed adequately
When sawing, move back and forth more and rinse
sufficiently
The saw blade cannot be
coupled
Coupling components or saw
blades deformed
Deformation
Have repaired by manufacturer
-
Incompatible saw blade
Insert a compatible Aesculap saw blade, see
Accessories/Spare parts
Saw blade cannot be
uncoupled
Locking pin not fully releasing
the saw blade
Push button for unlocking saw blade
not fully pushed
Fully press the saw blade release button, see
Attaching and detaching the saw blade
Issue
Cause
Detection
Remedy
Art. no.
Name
GA675
Lid
GA676
NiMH battery long
GA678
Sterile funnel
GA679
Battery removal aid
GB243R
GB244R
ECCOS screen basket with holder for Acculan 4
GB600
STERILIT Power Systems oil spray
GB495R
ECCOS fixation for machine
GB487R
ECCOS holder for lid
GB488R
ECCOS holder for sterile funnel
GB489R
ECCOS holder for battery removal aid
TA014542
Instructions for use for oscillating saw GA331 (A4
for ring binder)
TA014543
Instructions for use for oscillating saw GA331
(flyer)