Automatic cleaning/disinfection with manual pre-
cleaning
Note
The cleaning and disinfection device must fundamentally have a tested
efficacy (such as FDA approval or CE label pursuant to DIN EN ISO 15883).
Note
The cleaning and disinfection machine used for processing must be serviced
and checked at regular intervals.
7.8.1
Manual pre-cleaning with a brush
D–W:
Drinking water
RT:
Room temperature
►
Do not clean the product in a ultrasonic bath and do not immerse the
product in any fluids. Let any fluid incursions drain out immediately,
otherwise there is a danger of corrosion and loss of function.
Phase I
►
Mobilize non-rigid components during cleaning.
►
Thoroughly clean the product under running water.
Phase II
►
Mobilize non-rigid components during cleaning.
►
Brush difficult to access surfaces with a suitable plastic cleaning brush
for at least 1 min.
►
After manual preparation, check visible surfaces for residue and repeat
the pre-cleaning process as needed.
Note
For details on difficult to access surfaces, see Acculan pre-cleaning and
care information TA016000 (available in Aesculap Extranet at
https://extranet.bbraun.com).
Phase
Step
T
[°C/°F]
t
[min]
Conc.
[%]
Water
quality
Chemistry/Note
I
Rinsing
RT (cold)
–
–
D–W
until visually clean
II
Brushes
RT (cold)
–
–
D–W
until visually clean