![Azbil MagneW FLEX+ User Manual Download Page 80](http://html1.mh-extra.com/html/azbil/magnew-flex/magnew-flex_user-manual_3042665080.webp)
Installation
Azbil Corporation
2-54 MagneW
FLEX+/PLUS+
Electromagnetic Flowmeter Detector
(4)
Cut the outside shield off at a point 5 mm (0.2 inch) from the outside sheath,
leaving the specified length from the end of the outside shield.
CAUTION
Do not cut into the inside sheath.
5 mm
(0.2 inch)
Outside
shield
C-wire
5 mm
(0.2 inch)
White
C-wire
Black
Outside sheath
Inside sheath
Figure 2-50 Trimming the outside shield
(5)
Peel off the inside sheath so that a 15 mm (0.59 inch) section remains.
CAUTION
Do not cut into the inside shield.
15 mm
(0.59 inch)
C-wire
Inside
shield
White
C-wire
Black
15 mm
(0.59 inch)
Inside
sheath
Figure 2-51 Trimming the inside sheath
(6)
Cut the inside shield off at a point 5 mm (0.2 inch) from the edge.
5 mm
(0.2 inch)
Inside
shield
5 mm
(0.2 inch)
C-wire
A-wire
B-wire
C-wire
A-wire
B-wire
Inside
sheath
Inside
sheath
Figure 2-52 Attaching the lead wires (SA and SB)