Installation
MagneW
FLEX+/PLUS+
Electromagnetic Flowmeter Detector
2-49
5 mm
(0.2 inch)
Attach the lead wire
to the middle of this
5 mm (0.2 inch) length.
Outside
shield
C-wire
Lead wire
Outside
sheath
5 mm
(0.2 inch)
Figure 2-40 Attaching the C-wire
CAUTION
Be sure to use adequate solder so that contact failure will not occur between the
C-wire and the outside shield. If the soldering connection is insecure, noise is
generated due to contact resistance.
~Note
If the outside shield is not allowed to bend during soldering, in some
cases, the conductive tube cannot be peeled off due to the heat of the
solder.
(4)
Cut the outside shield off at a point 5 mm (0.2 inch) from the outside sheath,
leaving the specified length from the end of the outside shield.
CAUTION
Do not cut into the inside sheath.
5 mm
(0.2 inch)
Outside
shield
C-wire
5 mm
(0.2 inch)
White
C-wire
Black
Outside sheath
Inside sheath
Figure 2-41 Trimming the outside shield
(5)
Peel off the inside sheath so that a 15 mm (0.59 inch) section remains.
CAUTION
Do not cut into the inside shield.
~Note
Check the operating manual for the product you are using to
determine whether the lead wires (SA and SB) are necessary. If not,
skip to Step 8.