Subassembly Removal and Replacement
Hydrogen Peroxide Monitor Lamp Assembly (04-50544)
The hydrogen peroxide monitor lamp assembly includes the monitor
lamp optical window.
Lamp
Retainer
Screw
Ultratorr
Fitting
UV Lamp
Lamp
Housing
WARNING! ULTRAVIOLET (UV) LIGHT EXPOSURE!
ULTRAVIOLET LIGHT CAN HARM UNPROTECTED EYES
AND SKIN. DO NOT LOOK DIRECTLY AT AN ENERGIZED
UV LAMP.
HOT SURFACES! THE ULTRAVIOLET LAMP AND ITS
HOUSING ARE HOT WHEN ENERGIZED. CONTACT WITH
THESE PARTS MAY CAUSE BURNS OR OTHER INJURIES.
DO NOT TOUCH THE MONITOR LAMP ASSEMBLY WHEN
IT IS ENERGIZED.
CAUTION: Safe parts handling. Do not touch the surface of the
ultraviolet lamp or the optical window with your bare hands. Oil
and contaminants from your skin will damage these parts and
cause premature failure.
Required Tools:
1 mm hex wrench.
Soft-jaw pliers.
82 STERRAD
®
NX
™
Service Guide