Diagnostic Tests and Error Messages
4
Chapter 4.
Diagnostic Tests and
Error Messages
Diagnostic Files
Diagnostic files are created by the diagnostics tests. The files contain
details about the tests and the outcomes (Passed or Failed) of each. An
example of a Diagnostic file is shown in the figure following the table.
Touch
Diagnostic Files
to display a list of reports created by the
Diagnostics function. Scroll through the list and touch the report you
wish to view. Touch
View
to display the selected report. Touch
Back
to
return to the Additional Utilities menu.
Diagnostics
4
Note
: The temperature tests run automatically for about
1 hour. These tests should only be used if there is
a suspected temperature problem.
Touch
Diagnostics
to start automatic diagnostic testing of the sterilizer.
When started, the diagnostics function prompts you to select one of two
types of tests (either “Temperature” or “Other Tests”). If “Other Tests” is
selected, the sterilizer runs 10 operator-assisted tests of the sterilizer
subsystems. You may skip one or more tests in the automatic sequence
by touching
Cancel
when a test begins. The program advances to the
next test in the sequence.
The 10 tests and the sterilizer elements that are tested are listed in the
order in which they occur in the following table.
STERRAD
®
NX
™
Service Guide
39