Diagnostic Tests and Error Messages
4
DIAGNOSTICS REPORT
. . . . . . . . . . . . . .
File Name:
/xxxxxx
Power Supply Test
3.3 Volts Power Supply: x.x
5 Volts Power Supply: x.x
12 Volts Power Supply: xx.x
15 Volts Power Supply: xx.x
24 Volts Power Supply: xx.x
POWER SUPPLY TEST PASSED/FAILED
Time Stamp:
MM/DD/YY
HH:MM:SS
Pressure Test
PRESSURE TEST PASSED/FAILED
Time Stamp:
MM/DD/YY
HH:MM:SS
Plasma Test
PLASMA TEST PASSED/FAILED
Time Stamp:
MM/DD/YY
HH:MM:SS
Cassette Test
CASSETTE TEST PASSED/FAILED
Time Stamp:
MM/DD/YY
HH:MM:SS
H2O2 Sensor Test
H2O2 SENSOR TEST PASSED/FAILED
Time Stamp:
MM/DD/YY
HH:MM:SS
Display Test
DISPLAY TEST PASSED/FAILED
Time Stamp:
MM/DD/YY
HH:MM:SS
Printer Test
PRINTER TEST PASSED/FAILED
Time Stamp:
MM/DD/YY
HH:MM:SS
Fan Test
FAN TEST PASSED/FAILED
Time Stamp:
MM/DD/YY
HH:MM:SS
Sound Test
SOUND TEST PASSED/FAILED
Time Stamp:
MM/DD/YY
HH:MM:SS
Character Set
. . . . . . . . . . . . . .
ABCDEFGHIJKLMNOPQRSTUVWXZY
a b c d e f g h i j k l m n o p q r s t u v w x yz
! # $ % & ' ( ) * + , _ . / : ; < > ? @ [ ] ^ _ { | }
0123456789
Display Test
DISPLAY TEST PASSED/FAILED
Time Stamp:
MM/DD/YY
HH:MM:SS
Temperature Test
Door Rate:
x.x c/min
Chamber Low Rate:
x.x c/min
Chamber High Rate:
x.x c/min
Vaporizer Rate:
x.x c/min
Condenser Heat Rate:
x.x c/min
Condenser Cool Rate:
x.x c/min
Figure 7. Diagnostic File Example
STERRAD
®
NX
™
Service Guide
41
Summary of Contents for sterrad NX
Page 1: ...STERRAD NX Sterilization System Service Troubleshooting Guide...
Page 10: ......
Page 16: ......
Page 21: ...Functional Description 3 Figure 1 Software Control Diagram STERRAD NX Service Guide 21...
Page 38: ......
Page 100: ......
Page 124: ......