5
Subassembly Removal and Replacement
Latch Assembly (74-51398)
The alignment of the door and the latch must be performed by adjusting
screws located in the top and bottom door hinges. Refer to the subsection
titled
Door Assembly
for details.
Figure 29. Latch Assembly
Required Tools:
3 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel, front bottom panel, top panel, and
right side panel.
Removal:
1. Unplug cable P116 from the door open/closed sensor.
2. Disconnect the two slide-on connectors to the door open/closed
sensor.
3. Remove two screws that attach the sensor and bracket to the
chassis.
4. Pull the sensor away from the chassis to remove.
5. Remove two screws that attach the keeper mechanism to the
interior surface of the door.
6. Pull the keeper mechanism off the door.
Replacement:
Replacement is the reverse of the removal steps.
88 STERRAD
®
NX
™
Service Guide