Maintenance
6
Condenser/Vaporizer Thermistors Replacement
1. Slide the blanket aside to access the thermistors.
2. Unscrew the thermistors.
To install the new thermistors, do the following:
1. Apply a small amount of thermal compound to the bottom of the
thermistor lugs.
2. Attach the thermistors in locations shown in the following figure.
Figure 41. Condenser/Vaporizer Thermistors.
Cassette Disposal and Printer
Maintenance
The procedures in this section are routinely performed by the customer.
If the cassette box becomes full (3 cassettes) during a planned
maintenance procedure, it is a good idea to dispose of the empty
cassettes and insert a new box in the collection drawer.
IMPORTANT! You
must
use the cassette disposal box to dispose of
empty cassettes. Never reuse a cassette disposal box. Once a cassette
disposal box has been removed from the drawer, a new cassette disposal
box must be assembled and inserted in the drawer.
4
Note
: Cassettes with unused hydrogen peroxide are
hazardous waste as defined by the EPA and should
be disposed of accordingly.