Functional
Description
3
Process Variables and Cancellation
Limits
The following table lists the control parameters and cancellation limits
for the Standard and Advanced cycles.
Variable
Standard Cycle
Advanced Cycle
Cycle time
28 minutes
38 minutes
Chamber wall temperature
50º C
50º C
Pressure at delivery to vaporizer
Atmospheric pressure
Atmospheric pressure
Chamber pressure before transfer to
chamber.
≤
300 mtorr
≤
300 mtorr
Transfer time
3 minutes
7 minutes
Diffusion time
0.5 min
0.5 min
Plasma time
4 min
4 min
Chamber pressure
800 mtorr
800 mtorr
Plasma power
500 Watts
500 Watts
STERRAD
®
NX
™
Service Guide
37