Subassembly Removal and Replacement
5
Oil Mist Filter (25-51970)
Figure 33. Oil Mist Filter
This is a PM procedure.
Required Tools:
4 mm hex wrench.
6 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel, top panel, front bottom panel, and
right side panel.
3. Remove the vacuum pump assembly.
Removal:
1. Unscrew the 4 screws fastening the housing.
2. Separate the housing and the filter.
3. Remove the filter unit.
Replacement:
Replacement is the reverse of the removal procedure.