5
Subassembly Removal and Replacement
Display Assembly (40-50532)
Although they are separate parts, the touch-screen display and the
backlight are replaced together whenever a failure occurs in one.
Screw
Screw
Screw
Screw
Figure 22. Display Assembly
WARNING! ELECTRIC SHOCK HAZARD. HAZARDOUS
VOLTAGES ARE PRESENT INSIDE THE DISPLAY
BACKLIGHT. TURN THE STERILIZER OFF BEFORE
HANDLING THE DISPLAY ASSEMBLY.
Required Tools:
3 mm hex wrench.
7 mm nut driver.
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel and the top panel.
Removal:
1. Unplug 2 cables: P21 and P29.
2. Loosen the 2 screws that attach the door lock assembly to the
chassis and move the assembly to expose the display bracket nuts.
76 STERRAD
®
NX
™
Service Guide
Summary of Contents for sterrad NX
Page 1: ...STERRAD NX Sterilization System Service Troubleshooting Guide...
Page 10: ......
Page 16: ......
Page 21: ...Functional Description 3 Figure 1 Software Control Diagram STERRAD NX Service Guide 21...
Page 38: ......
Page 100: ......
Page 124: ......