Subassembly Removal and Replacement
5
Ultraviolet Lamp Power Supply (40-51449)
The ultraviolet (UV) lamp is part of the hydrogen peroxide monitor
system. The power supply is located in the top module and removal and
replacement procedures are described here. The hydrogen peroxide
monitor lamp assembly and the hydrogen peroxide detector assembly are
described in the “Chamber Module” section of this chapter.
Figure 26. UV Lamp Power Supply
WARNING! ELECTRIC SHOCK HAZARD. HAZARDOUS
VOLTAGES ARE PRESENT INSIDE THE UV LAMP POWER
SUPPLY. TURN OFF THE LAMP USING THE TEST SCREEN.
THEN TURN THE STERILIZER OFF BEFORE HANDLING
THE POWER SUPPLY.
Required Tools:
3 mm hex wrench.
Preparation:
1. Navigate to the test screen.
2. Turn the lamp OFF.
3. Turn the sterilizer OFF.
4. Remove the front top panel and the top panel.
Removal:
1. Unplug cables: P20 and P120.
2. Remove the two screws that attach the power supply to the
chassis.
3. Lift the power supply to remove.
Replacement:
Replacement is the reverse of the removal steps.
STERRAD
®
NX
™
Service Guide