5
Subassembly Removal and Replacement
Right Side Panel (33-51193)
Screw
Screw
Figure 12. Right Side Panel
Required Tools:
3 mm hex wrench.
Preparation:
1. Turn the sterilizer OFF.
2. Remove the front top panel, the top panel, and the front bottom
panel.
Removal:
1. Remove the two screws on the top of the right panel.
2. Pull the top of the panel away from the chassis and lift the panel
to remove.
Replacement:
Replacement is the reverse of the removal steps.