TD 93021US
17 July 2017 / Ver. PF3
69
Installation Guide
teleCARE IP
Figure 91. Bedside module NIMB2 electrical connections and address setting
The address is selected using the DIP switch on the switch module printed circuit board. For details of the
DIP switch settings see
6.4.10, Dip Switch Settings on page 67.
For details of the 4-Pole connector see
6.4.1, 4-Pole Connector Terminal on page 60
For light switching relay details see
6.5.2, Light Switching Relay Maximum Load and Surge Damping
.
For details of
connecting a speech module see
6.15.1, Connecting the Speech Module on page 101
For details of
connecting the external call see
6.5.3, External Call Contact with Call Lamp Output on page
Note:
The 4-pole connector terminals required for the room bus and other inputs and
outputs are not supplied with the switch module. They are available as accessories and
must be ordered separately.
Not supported
with UL 2560
systems
4-Pole Connector
for the
Room Bus
Dip switch
for setting the
Room Bus
Address
0, 1, 2, or 3
4-Pole Connector
for an External Call Input
and Call LED Output
Connector for
the Speech
Module
4-Pole Connector for the
Stereo Input from the
Television Interface
4-Pole Connector for
the Light Switching
Relays