Installation Guide
teleCARE IP
6.4.10 Dip Switch Settings
Each room bus provides eight addresses for the connection of teleCARE IP switch modules. The lower
four of these addresses are set by a DIP switch that is located on the PCB of the certain switch module.
The upper four addresses cannot be set as they are dedicated to certain switch modules which have the
appropriate address permanently set in the switch module.
The image below shows the location of the DIP switch which is in the same place on all active switch
modules.
Figure 89. Active switch module DIP switch location
The the DIP switch allows 4 addresses to be set: 0, 1, 2 and 3. These settings are only used for active
switch modules and active socket modules.
Address 4 is used for the toilet cancel module and this module has the address fixed and it is not set by a
DIP switch.
The addresses 5 to 7 are used for special teleCARE IP peripherals consisting of the corridor lamp, duty
selector, card reader and room display. These module have the address fixed and it is not set by a DIP
switch.
The room bus address assignments are summarized in the following table:
Room Bus Address
Active Peripheral
Address Setting
0 - 3
Door side Modules
Bedside Modules
Pull Cord Modules
Medical Rail Socket
Set by DIP switch
4
Toilet Cancel Module
Fixed
5
Slave Corridor Lamp
Duty Selector
Fixed
6
Card reader
Fixed
6 + 7
Room Display
Fixed
DIP Switch
Settings:
0, 1, 2, 3