TD 93021US
17 July 2017 / Ver. PF3
81
Installation Guide
teleCARE IP
Note:
The 4-pole connector terminals required for the room bus and the passive bus must
be ordered separately as accessories.
6.10 Pull Cord Module - Passive (NIPC-XXP)
Figure 107. Passive pull cord module 3 buttons
The Passive Pull Cord Module is intended for use in the teleCARE IP system, in areas such as bath-rooms
and toilets. It must be connected to the passive bus of a door side module or active pull cord cancel
module. When connected to the door side module, it can be combined with the passive toilet cancel
module on the same passive bus. The passive bus consists of four wires (see
).
Note:
Only one Passive Pull Cord Module can be connected per passive bus.
For assembly and attaching instructions for the pull cord refer to
6.8.2, Assembling and Attaching the Pull
.
The Passive Pull Cord Module requires a single backplate which must be ordered separately. The
backplate enables this switch module to be mounted over different types of back boxes. Alternatively, a
spacer with installation kit is available for surface mounting.
6.10.1 Pull Cord Module (Passive) Electrical Connections
The passive pull cord module is a “passive peripheral” therefore it must be connected to the passive bus
of the linked door side module or active pull cord module by the 4-pole passive bus connector (see