TD 93021US
17 July 2017 / Ver. PF3
19
Installation Guide
teleCARE IP
5
Control Equipment
This section describes the installation instructions for the following products:
•
“Room Controller (NIRC3)” (details on page 19)
•
“Blank Front Cover for the Room Controller” (details on page 21)
•
“Voice Piggy Back (NIVP)” (details on page 29)
•
“Corridor Lamp (NICL2)” (details on page 43)
•
“System Manager (NISM2)” (details on page 51)
5.1
Preparation
It is important to refer to the following teleCARE IP control module installation instructions for complete
electrical connection and assembly details before starting the installation.
The Ingress Protection of the control modules is IP40, therefore the areas in which the teleCARE
equipment is to be installed must be clean, dry, and weatherproof.
The walls on which the control modules are to be installed should be finished (painted, wall papered etc.)
before the control module is installed.
It is important to have the appropriate length of free cable pulled out at every location where the control
module is to be installed. The length of stripped cable and length of exposed copper wire must conform
to the relevant installation instructions. Electrical power to the equipment must be off before connecting
any devices.
5.2
Room Controller (NIRC3)
The Room Controller (NIRC3) has a translucent dome cover which accepts up to four LED boards which
must be ordered separately. The LEDs are used for the signaling of calls, attendant presence and faults.
The LED board is available in five colors: red, green, yellow, white and blue.
Figure 11. The teleCARE IP room controller with corridor lamp