Troubleshooting
5-10
Angel
®
cPRP System
Operator’s Manual
10.
Sensor Alarm won’t clear
a. Caused by a dirty sensor or light source too close to the sensor
i. Clean sensor with lens cloth or similar material
ii. Ensure that the tubing/cuvette does not contain residual blood upon start up
iii. Move device (or light source) to prevent illumination issue. Light sources should be at
least 3 feet from the Angel sensor system.
11.
Valve assembly won’t attach to key appropriately
a. Occurs when “end-case” was not selected after the previous processing procedure
i. Start “new case” or manually turn valve manifold on the processing set disposable to
match the position of the valve attachment key on the Angel machine.
12.
The pump rotor spins intermittently and won’t draw blood
a. Occurs when peristaltic pump head is not properly installed (see illustrations below for details).
i. Remove tubing (turn the rotor counter clockwise if needed),
ii. Align rotor head on spindle,
iii. Push down on the Rotor and turn clockwise,
iv. Reinstall tubing (turn the rotor clockwise) and continue case.
If you continue to experience difficulty, contact Arthrex, Inc. Customer Service
Summary of Contents for Angel
Page 2: ...This page intentionally left blank ...
Page 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 14: ...Overview xii Angel cPRP System Operator s Manual ...
Page 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Page 18: ......
Page 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...