Routine Care
Angel
®
cPRP System
Operator’s Manual
6-5
“Download software from this source” Button
Touching the “
Download software from this source
” Button
begins the installation process for the selected software version.
When the button is touched, the selected software version is copied to the
Angel System
’s
temporary memory and is validated. It is not installed, until the operator confirms the installation
from the Install Screen. Refer to “
New Software
” on page 6-1 for details.
“Re-check sources” Button
Touching the “
Re-check sources
” Button will cause the
Angel System
to
check for any storage device connected to the USB Port for valid software to
install. This is used if the source was not found, or was not properly connected when the
“
Check for new software
” Button was touched
Visual Inspection
The
Angel System
should be inspected periodically for any problems such as bent or broken switches,
frayed or twisted power cords, and loose or missing hardware. If the device displays any of the above
conditions, discontinue use of the
Angel System
until the problem is corrected and it has been verified
that the device is operating correctly.
Verify that the USB Port located on the exterior of the
Angel System
, is intact and not damaged.
Verify that the Stator Arm is not loose on its pivot pin. It should only rotate up and down with hand
pressure, and have no side-to-side motion. If the arm is loose, lift it to the raised braking position and
locate the cap screw near the base of the Stator Arm. Using a 9/64 inch hex wrench, tighten the cap
screw against the pivot pin.
Routine Care
External Surfaces
Angel System
are to be thoroughly cleaned and disinfected according to the following instructions..
Note:
In the event of minor fluid (blood) spills that only affect the exterior of the
Angel System
(and do
not accumulate in the Centrifuge Well), such spills should be addressed
within 10 minutes
of the
occurrence, if possible, and should not be allowed to dry on the exterior surfaces of the unit. For minor
fluid (blood) spills, ensure that the pump rotor is removed for cleaning and disinfection processes per
the following.
1. Remove the Pump Rotor from the rotor housing by pushing down and turning the rotor counter-
clockwise.
Summary of Contents for Angel
Page 2: ...This page intentionally left blank ...
Page 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 14: ...Overview xii Angel cPRP System Operator s Manual ...
Page 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Page 18: ......
Page 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...