Technical Data
7-2
Angel
®
cPRP System
Operator’s Manual
Characteristic
Performance
Condition
Operating Atmospheric
Pressure
70 kPa to 106 kPa
Storage Temperature
-20° to 37.7°C
(-4° to 100°F)
Storage Humidity
10 to 90%
Non-condensing
Storage Atmospheric
Pressure
50 kPa to 106 kPa
Ingress Protection
IPX0
Fuse Rating
2x F 5A L, 250V
Restrictions
Not to be used in an
explosive atmosphere
Guidance and manufacturer’s declaration – electromagnetic emissions
The
Angel System
is intended for use in a professional healthcare facility environment, details for
the electromagnetic environment are specified below. The operator of the
Angel System
should
assure that it is used in this environment. If the
Angel System
is used within these prescribed
environments, the system will continue to process whole blood or a mixture of whole blood and
bone marrow without manufacturer service.
Emission test
Compliance Level
Electromagnetic
environment -
guidance
RF emissions CISPR 11
used for domestic
Group 1
The
Angel System
uses
RF energy only for its
internal function.
Therefore, its RF
emissions are very low
and are not likely to
cause any interference
in nearby equipment.
RF emissions CISPR 11
A
The
Angel System
is
suitable for use in all
establishments other
than domestic and
those directly connected
to the public low-voltage
power supply network
that supplies buildings.
Harmonic emissions
IEC 61000-3-2
A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
Warning:
The emissions characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 Class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might not offer adequate protection to
radio-frequency communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
Warning:
Use of the
Angel System
adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
Summary of Contents for Angel
Page 2: ...This page intentionally left blank ...
Page 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 14: ...Overview xii Angel cPRP System Operator s Manual ...
Page 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Page 18: ......
Page 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...