Troubleshooting
5-2
Angel
®
cPRP System
Operator’s Manual
is pressed, the centrifuge is
stopped and the “
Lid Latch
”
alarm is sounded.
Inc. Customer Service.
Lid open
yellow
red
If the lid is open
before
the
Start Button is pressed,
the
yellow
icon appears,
prompting you to close the
lid properly.
If the lid opens
during
the cycle,
the
red
icon appears. The
centrifuge stops until the
problem is corrected.
Close the lid cover. Ensure
the tubing is not caught
between the lid and body of
the machine.
Light sensor
calibration
The “
Light sensor
calibration
” alarm is
displayed if the Platelet
Sensor is not able to
calibrate itself. The three
reasons this may occur are:
(1) the user has not loaded
the disposable set correctly,
(2) the Platelet Sensor is
dirty, or
(3) the Platelet Sensor is too
close to an external light
source.
(1) If the disposable set is not
yet installed, install it
according to the
instructions in
Chapter 2:
Installing the
Disposables
. Make sure
that the Platelet Cuvette is
securely in place, and that
the Valve Assembly is
properly mounted.
(2) If the alarm is still active,
clean the Platelet Sensor,
then reinstall the
disposable set. (Please
see the “Note” on 6-8)
(3) Ensure that the Platelet
Sensor is at least 1 meter
(3 feet) from all external
light sources.
Machine
temperature
(Machine temp)
If the
Angel System
detects
that the temperature is
unacceptably high, the
“
Machine temperature
”
alarm is sounded.
Power down the
Angel System
for a short time before any
further processing. If this alarm
continues, contact Arthrex, Inc.
Customer Service.
Plate locking
error
If the Separation Chamber
Plate is improperly attached
to the Centrifuge Adapter,
the “
Plate locking
” alarm
is sounded.
Turn separation chamber
clockwise until position
indicator snaps into place.
Touch the Movie icon to
see an animation which
demonstrates how to properly
load and lock the Separation
Chamber Plate.
PPP bag full
If more than 300 mL of fluid has
been pumped into the PPP
Compartment of the Three-
Compartment Reservoir Bag,
the “
PPP bag full
” alarm
is sounded.
Empty the PPP Compartment
of the Three-Compartment
Reservoir Bag, and then
continue processing.
Summary of Contents for Angel
Page 2: ...This page intentionally left blank ...
Page 8: ...Overview vi Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 14: ...Overview xii Angel cPRP System Operator s Manual ...
Page 17: ...Overview Angel cPRP System Operator s Manual xv This page intentionally left blank ...
Page 18: ......
Page 26: ...Overview 1 8 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 66: ...Processing 3 28 Angel cPRP System Operator s Manual This page intentionally left blank ...
Page 98: ...Routine Care 6 12 Angel cPRP System Operator s Manual This page intentionally left blank ...