1. Remove the six (6) locking cap screws, in the following way (see Figure 6-7 below). First completely
remove the cap screws 1-4 (Important: note the removal pattern), and then alternate between
screws 5 and 6, two turns at a time. Set aside the screws for cleaning.
1. Spring Plate
2. Adapter Top
Figure 6-7 Removing the Centrifuge Adaptor Plate
2. Remove the spring plate, taking care not to misplace the springs residing beneath the spring plate.
Remove the springs and set aside for cleaning. Remove the Centrifuge Plate by removing the four
cap screws that secure the Centrifuge Plate to the motor spindle.
1. Springs (x8)
2. Centrifuge Plate
3. Spindle