5-27
The
Hold To Start
display will appear temporarily, as
shown at right.
The
Release To Start
display will appear, as shown at
right.
2.
Release
Change Equip
button. The
Add / Remove
Equipment
display
will appear, as shown at right.
3.
Press
Change Equip
button. The
Do Not Touch Bed
display
will appear temporarily, as shown at right. Do
not touch unit until patient weight returns in
Scale
display.
Bed Motion Control Panel
Use the bed motion control panel to adjust the height and angle of the patient surface and frame and lock out
frame functions. The bed motion control panel is located on the footboard below the main control panel.
Height of patient surface will automatically adjust as needed to accommodate Trendelenburg and
Reverse Trendelenburg.
HOLD TO START
RELEASE TO START
ADD / REMOVE EQUIPMENT
DO NOT TOUCH BED
SIDE RAIL CONTROL LIGHTS
Press to sequentially turn patient control
panel backlight on low, medium, high or off.
BED UP / DOWN BUTTONS
Use to adjust height of hard-pan surface
from 22.50 in to 35 in (57 cm to 89 cm).
BED POSITIONING BUTTONS
Use to adjust head up / down
and knee gatch up / down
BED MOTION LOCK
Press to lock-out all bed
motion functions.
TRENDELENBURG / REVERSE
TRENDELENBURG
Use to adjust patient support surface
from +9º angle Trendelenburg to -9º
angle Trendelenburg.
SIDE RAIL CONTROL LOCK-OUTS
Press to lock out side rail control for
head up / down, knee up / down and
bed up / down functions
SIDERAIL
CONTROL
LIGHTS
BED
MOTION
LOCK
SIDERAIL CONTROL
LOCKOUTS
BED MOTION
Trend.
BACK
KNEE
ON
OFF
OFF
ON
BACK
KNEE
UP
DOWN
UP
DOWN
UP
UP
DOWN
DOWN